Technical data
Dimensions (W/D/H)
|
|
Unit |
Value |
|---|---|---|
|
Footprint (approx.) |
mm in |
260 x 260 10.2 x 10.2 |
|
Housing fully closed (approx.) |
mm in |
360 x 370 x 490 14.2 x 14.6 x 19.3 |
|
Minimum required space for operation (approx.) |
mm in |
560 x 620 x 530 15 x 25 x 21 |
Base system
|
|
Unit |
Value |
|---|---|---|
|
Weight (approx.) |
kg lbs |
29 64 |
|
Axes – Rotational axis A |
|
4 +190° to -10° |
|
Overvoltage category (IEC 60664-1) |
|
II |
External air extraction system
|
|
Unit |
Value |
|---|---|---|
|
Minimum extraction capacity |
l/min cfm |
2,500 (at 220 hPa) 88.3 (at 3.2 psi) |
|
Suction unit |
|
Designed for the commercial use in the dental sector Equipped with a filter of the filter class M |
Ambient conditions
|
|
Unit |
Value |
|---|---|---|
|
Relative humidity |
|
80 %, non-condensing |
|
Ambient temperature for storage/transport |
°C °F |
‑20 – 60 -4 – 140 |
|
Ambient temperature for operation |
°C °F |
10 – 35 50 – 95 |
|
Location type |
|
Indoor |
|
Maximum height above mean sea level |
m ft |
2000 6561 |
|
Ambient air (IEC 60664-1) |
|
Dust free, pollution degree 2 |
|
Electromagnetic radiation (IEC 61326-1) |
|
Residential, business and commercial areas |
|
Electromagnetic interference (IEC 61326-1) |
|
Industrial environment |
Sound emission
|
Operating condition |
Unit |
Value |
|---|---|---|
|
Machining |
dB(A) |
74.55 |
|
All other operating conditions |
dB(A) |
<70 |
|
Measuring conditions |
Value |
|---|---|
|
Machined material |
Composite (VITA CAD temp block) |
|
Condition of the tools |
worn |
|
Measured value |
Sound pressure level (distance: 1 m) |
|
Measurement according to |
ISO 3746, accuracy grade 3 |